Certain System 83 Plus 2 and System 83 Plus 9 units running the Windows 7 operating system were not fully validated prior to distribution.
Source facts
Action
A notification letter, dated February 26, 2013, was sent to direct accounts to inform them of the recall and that technicians would visit facilities to replace the Windows 7 operating system with an appropriately validated Windows XP operating system.
Initiated
2013-03-12
Root cause
Software design
Terminated
2014-06-26
Distribution
Worldwide Distribution: USA (nationwide) and Internationally to: Canada, Slovakia Republic, Sydney and Peru.
Recall number
Z-0426-2014
Recall status
Terminated
CDRH record ID
121143
FDA classified
2013-11-27
Recall event number
65985