Skip to content

Public FDA record

Z-0426-2021: EDI Novel Coronavirus COVID-19 IgG ELISA Kit, REF: KT-1032 Per label: This kit is intended for qualitat

Recall record Z-0426-2021 for Epitope Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0426-2021
Company in FDA record
Epitope Diagnostics, Inc.
FDA record creation date
Nov 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The use of deionized water rather than distilled water during the manufacturing process, may have resulted in accelerated decreased OD signals for the Positive Control value.

Source facts

Action
On 10/01/20, the firm, Epitope Diagnostics Inc. (EDI), sent "URGENT: EDI" Novel Coronavirus COVID-19 IgG ELISA Kit Recall" notices were sent to customers. Customers were asked to do the following: Results where the assay controls do not meet the manufacturer's defined specifications should be considered invalid and repeated with another lot. Immediately examine your inventory and quarantine product subject to recall. In addition, if you have further distributed this product, please identify your customers and notify them at once of this product recall. Complete and promptly return the Customer Response Form by Email to: cs@epitopediagnostics.com or Fax to: 858-693-7877. Your notification to your customers may be enhanced by including a copy of this recall notification letter. For customers located within United States, we will ask that you return this affected product to the recalling firm. For customers located outside United States, please destroy the affected product. For any questions, please refer to EDIs Quality Assurance and Regulatory Affairs Manager, Stefanie Dallezotte, by email: lenart@epitopediagnostics.com or by telephone: +1 858 693 7877.
Initiated
2020-10-01
Root cause
Nonconforming Material/Component
Distribution
U.S Distribution.: NH, CO, OR, CA, FL, NY, MD, NJ, IL, NC, MI, TX, PA, MA, OH, NM, VA, SC, HI, and TN.
Recall number
Z-0426-2021
Recall status
Open, Classified
CDRH record ID
183989
FDA record created
2020-11-06
Recall event number
86516

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗