Skip to content

Public FDA record

Z-0427-2014: Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Produc

Recall record Z-0427-2014 for Cutera, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0427-2014
Company in FDA record
Cutera, Inc.
FDA classification date
Dec 9, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cutera received two adverse events associated with the edge of the electrode, when it was pressed against neck skin over the thyroid cartilage and due to the absence of adequate underlying conductive tissue, increased skin heating occurred. Both patients developed linear welts and blisters consistent with partial thickness thermal burns to the skin.

Source facts

Action
Cutera sent an IMPORTANT SAFETY INFORMATION letter dated October 16, 2013, to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. It is important that you share this correspondence and the updated Operator Manual andvtreatment guidelines with all staff at your office that use the truSculpt, so they are aware of this new safety information. Please immediately remove and dispose of the current truSculpt Operator Manual and clinical guidelines and replace them with the new and revised truSculpt Operator Manual and clinical guidelines. Customers with questions were instructed to call 415-657-5500 or 415-657-5568.
Initiated
2013-10-16
Root cause
Employee error
Terminated
2013-12-09
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada
Recall number
Z-0427-2014
Recall status
Terminated
CDRH record ID
122929
FDA classified
2013-12-09
Recall event number
66674

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗