Skip to content

Public FDA record

Z-0428-2021: QUIDEL Sofia SARS Antigen FIA IVD REF 20374

Recall record Z-0428-2021 for Quidel Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0428-2021
Company in FDA record
Quidel Corporation
FDA record creation date
Nov 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to false positive results.

Source facts

Action
On 10/14/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification via US Certified mail to customer informing them that a limited number of SARS Antigen FIA test cassettes have an increased potential for false positive results. Customers are instructed to stop using all product from the affected lot and return the unused product to the Recalling Firm for a replacement or issuance of credit fro any unused product. The specific instruction to customers are: 1. Stop using Sofia SARS Antigen FIA, Cat. #20374, Lot 706142. 2. Return the attached Verification Form with the amount of product that will be returned so a credit may be issued. The customer will be contacted to arrange the return of product and the corresponding credit. 3. Return any unused product from Sofia SARS Antigen FIA, Cat. #20374, Lot 706142. 4. In the event that the customer's facility has used this lot number to report patient results, please contact Technical Support at customernotifications@quidel.com. Technical Support will conduct an analyzer log data review to determine if any reported patient results were impacted.
Initiated
2020-09-25
Root cause
Process control
Terminated
2024-09-18
Distribution
U.S.: AL, AR, AZ, CA, FL, GA, IN, LA, MO, MS, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WY
Recall number
Z-0428-2021
Recall status
Terminated
CDRH record ID
184011
FDA record created
2020-11-06
Recall event number
86550

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗