Skip to content

Public FDA record

Z-0429-2013: Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High).

Recall record Z-0429-2013 for Horiba Instruments, Inc dba Horiba Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0429-2013
Company in FDA record
Horiba Instruments, Inc dba Horiba Medical
FDA classification date
Nov 21, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class III

Reason stated in the source

HORIBA Medical Irvine Technical Support received reports from some customers that have experienced high recover values for Hemoglobin when using Minotrol 16 lot #MX090 (Level High). Customer communication was sent 10/01/2010.

Source facts

Action
HORIBA Medical sent a Product Information Notice letter dated November 5, 2010 via Fax. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to follow the instructions included with the Product Information Notice. For questions customers were instructed to call the Technical Support Hotline at 1-888-903-5001. For questions regarding this recall call 949-453-0500, ext 208.
Initiated
2012-09-14
Root cause
Nonconforming Material/Component
Terminated
2013-04-22
Distribution
Nationwide Distribution
Recall number
Z-0429-2013
Recall status
Terminated
CDRH record ID
114174
FDA classified
2012-11-21
Recall event number
63539

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗