A calibrator for an in vitro diagnostic kit is not standarized correctly and may cause a calibration error when measuring microbiological cultures for turbidity.
Source facts
Action
The recalling firm notified consignees by phone, fax and e-mail on 12/01/05. The notification advised that nephelometers displaying 'CAL' error messages on the LED should not be used until calibrated with a different kit lot. Consignees were requested to discard referenced lots for replacement. A response form was included in the notification.
Initiated
2005-12-01
Root cause
Other
Terminated
2006-04-06
Distribution
Product was sold to direct accounts nationwide and international affiliated distributors in Japan, Belize, Taiwan, Canada, Korea, Australia, and the Philippines.
Recall number
Z-0430-06
Recall status
Terminated
CDRH record ID
43266
FDA classified
2006-01-24
Recall event number
34202