Open original FDA record ↗Recall record
- FDA identifier
- Z-0432-2013
- Company in FDA record
- OrthoPediatrics Corp
- FDA classification date
- Nov 22, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Complaint received from sales representative that the drill guide tubes would not screw into replenishment instrument. Upon investigation, it has been determined that the device was not manufactured to design specifications and will not function as intended.
Source facts
- Action
- OrthoPediatrics sent an "URGENT RECALL" e-mail of the Trochar Scalpel Instrument to all affected customers. The e-mail identifies the product, problem, and actions to be taken by the customers. The e-mail states that the instrument has been manufactured incorrectly and will not work (function as intended). The threads in the instrument are faulty and the guide tubes will not thread in correctly. Customers will receive follow up information.
- Initiated
- 2012-10-18
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-09-18
- Distribution
- Nationwide Distribution-including the states of OH, NC, CO, FL, IN, MI, and TX.
- Recall number
- Z-0432-2013
- Recall status
- Terminated
- CDRH record ID
- 114199
- FDA classified
- 2012-11-22
- Recall event number
- 63554