Z-0432-2021: FlexLab (FLX) Automation System with the Aliquoter Module (Inpeco P/N FLX-212) The FlexLab (aka Accelera

Recall record Z-0432-2021 for Inpeco S.A.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0432-2021
Company in FDA record
Inpeco S.A.
FDA record creation date
Nov 6, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.

Source facts

Action
The firm, inpeco, sent "Urgent: Medical Device Correction" letters dated 7/14/20 to its customers. The letter described the product, problem and actions to be taken. Action to be taken by the user: Be aware that the Secondary Sample Tube flagged with error 2132 or 1442 may be diluted: discard this Secondary Sample Tube or manage it according to your laboratory guidelines. Your service provider will contact you to schedule the upgrade of the software driver and language.ini with the new error message and error recovery displayed if error 2132 or 1442 occurs. Until the service visit please maintain awareness on this notice. Please transfer this notice to whom it might concern. Please complete and return the Field Safety Notice Receipt Confirmation form attached to this letter within 15 days directly to the email address specified in the email communication. Contact reference person: For any clarification you may need, do not hesitate to contact: Eva Balzarotti - Regulatory Affairs Manager E-mail: Regulatory.Affairs@inpeco.com Phone: (+41) 919118 224
Initiated
2020-07-14
Root cause
Software design
Terminated
2023-12-28
Distribution
Global Distribution including: USA and countries of: Algeria, Australia, Azerbaijan, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Egypt, Finland, France, Germany, Indonesia, Israel, Italy, Libano, Mexico, Netherlands, Philippines, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom, and United Arab Emirates.
Recall number
Z-0432-2021
Recall status
Terminated
CDRH record ID
184082
FDA record created
2020-11-06
Recall event number
86581

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗