Open original FDA record ↗Recall record
- FDA identifier
- Z-0433-2014
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Dec 3, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System that should have been held for manual review due to auto-verification rules or the delay/omission of result transmission to the LIS.
Source facts
- Action
- Siemens sent an Urgent Medical Device Correction letter in May 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the letter in its entirety and ensure that all operators understand the information presented. Specific actions needed to be implemented by the customer were outlined in the letter. Customers were instructed to retain the letter with their laboratory records, and forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Technical Solutions Center or their local Siemens technical support representative. For questions regarding this recall call 800-441-9250.
- Initiated
- 2013-05-10
- Root cause
- Software design
- Terminated
- 2014-08-26
- Distribution
- Worldwide Distribution: USA (nationwide) and internationally to Canada.
- Recall number
- Z-0433-2014
- Recall status
- Terminated
- CDRH record ID
- 118015
- FDA classified
- 2013-12-03
- Recall event number
- 65147