Z-0433-2026: Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Versio

Recall record Z-0433-2026 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0433-2026
Company in FDA record
Elekta, Inc.
FDA classification date
Nov 5, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound

Source facts

Action
On 09/30/2025, the firm sent via email an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that specified serial numbers, there is an increased risk that paint flakes from the engravings may separate from the Leksell Vantage Arc System components. Separation of paint flakes is likely to occur during cleaning and sterilization. If paint flakes are not washed away during the cleaning and sterilization process, these could remain on the Leksell Vantage Arc System when used in surgery. This poses a risk of paint flakes entering the surgical area. This situation has not been observed in clinical practice, but the risk should be fully considered and mitigated.
Initiated
2025-09-30
Root cause
Process change control
Distribution
Worldwide - U.S. Nationwide distribution in the states of AR, CA, CO, FL, ID, IL, LA, MA, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, WA, WI, and WV. The countries of Algeria, Australia, Austria, Belgium, China, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Republic of Korea, Kuwait, Lithuania, Mexico, Netherlands, New Zealand, Norway, Pakistan, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, and United Kingdom.
Recall number
Z-0433-2026
Recall status
Open, Classified
CDRH record ID
216372
FDA classified
2025-11-05
Recall event number
97705

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗