Skip to content

Public FDA record

Z-0434-05: VISTA BRITE TIP RDC(1) Guiding Catheter. Product Code 67021055, lot X0704002, Precutaneous Catheter.

Recall record Z-0434-05 for Cordis Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0434-05
Company in FDA record
Cordis Corporation
FDA classification date
Jan 26, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Two complaints have been received and confirmed, regarding a lot, indicating that the distal end had an 'RDC' shape rather than the 'RDC(1)' shape. The RDC shape has a larger curve. The product was manufactured with the incorrect forming wire during the guiding catheter manufacturing process.

Source facts

Action
Packets containing the Customer Letter and Acknowledgement Form for each domestic consignee were sent by overnight delivery to the appropriate Cordis Endovascular sales representatives. The letter is being hand-delivered to each initial domestic consignee by the Cordis Endovascular sales representative. Any remaining units at the account will be removed from the shelf and returned to Cordis. The account will be asked to sign the Acknowledgement Form.
Initiated
2004-12-07
Root cause
Other
Terminated
2008-11-19
Distribution
Distributed to 57 accounts throughout the United States and three international accounts.
Recall number
Z-0434-05
Recall status
Terminated
CDRH record ID
36253
FDA classified
2005-01-26
Recall event number
30637

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗