- FDA identifier
- Z-0434-05
- Company in FDA record
- Cordis Corporation
- FDA classification date
- Jan 26, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Two complaints have been received and confirmed, regarding a lot, indicating that the distal end had an 'RDC' shape rather than the 'RDC(1)' shape. The RDC shape has a larger curve. The product was manufactured with the incorrect forming wire during the guiding catheter manufacturing process.
Source facts
- Action
- Packets containing the Customer Letter and Acknowledgement Form for each domestic consignee were sent by overnight delivery to the appropriate Cordis Endovascular sales representatives. The letter is being hand-delivered to each initial domestic consignee by the Cordis Endovascular sales representative. Any remaining units at the account will be removed from the shelf and returned to Cordis. The account will be asked to sign the Acknowledgement Form.
- Initiated
- 2004-12-07
- Root cause
- Other
- Terminated
- 2008-11-19
- Distribution
- Distributed to 57 accounts throughout the United States and three international accounts.
- Recall number
- Z-0434-05
- Recall status
- Terminated
- CDRH record ID
- 36253
- FDA classified
- 2005-01-26
- Recall event number
- 30637