A manufacturing defect was observed for the membrane changer insert that does not allow electrolyte to be present under the new membrane. The PCO2 performance of a sensor without electrolyte will be deteriorated.
Source facts
Action
SenTec notified the one US distributor via e-mail on November 5, 2012. The distributor was advised to stop distribution of affected lot numbers. The communication to the distributor described the issue, identified affected product, and stated that SenTec will replace current stock at the end-customer level. The end user can discard the affected products. The distributor is asked to respond back with an e-mail to SenTec that they took appropriate action.
Initiated
2012-11-05
Root cause
Component design/selection
Terminated
2012-12-18
Distribution
Distributed only in MO.
Recall number
Z-0434-2013
Recall status
Terminated
CDRH record ID
114388
FDA classified
2012-11-22
Recall event number
63631