Skip to content

Public FDA record

Z-0434-2026: CEREPAK Detachable Coil System - Name/REF: UNIFORM 2MM X 2.5 CM/FCX100202, UNIFORM 2.5MM X 3.3 CM/FCX1002

Recall record Z-0434-2026 for Cerenovus Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0434-2026
Company in FDA record
Cerenovus Inc
FDA classification date
Nov 13, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Detachable coils system may fail to detach, which could result in hemorrhagic and ischemic stroke, procedural delays, or the need for additional surgical intervention.

Source facts

Action
On 10/2/2025, Johnson & Johnson Med Tech mailed "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" notices dated 10/1/2025 to customers who were asked to do the following: 1. Immediately stop using the affected devices. 2. Identify and segregate all affected devices that are in inventory / consignment at your facility. 3. Firm sales representative will coordinate the removal of the product from your facility and will coordinate the credit, as applicable, for any customer-owned inventory. 4. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 5. If any product subject to this recall has been forwarded to another facility, contact that facility, provide a copy of this letter and notify J&J sales representative. 6. If the product has been successfully implanted, there is no additional action needed. 7. Complete and return the attached Business Reply Form in Attachment 2 to: OneMD-Field-Actions@its.jnj.com. Please return the BRF even if you do not have product subject to this recall. 8. Keep this notice visibly posted for awareness until all product subject to this recall have been returned. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. If you have additional questions regarding this medical device recall (removal), please contact firm at OneMD-Field-Actions@its.jnj.com.
Initiated
2025-10-02
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution: CO, GA, MD, UT, OK, WI, MA, TN, SC, TX, CA, WV, LA, IL, NY, IA, AZ, AL, CT, WA, OH, ME, DC, FL, SD, KY, NJ, DE, NC, PA, MI, MO, VA, NH, NV and OR.
Recall number
Z-0434-2026
Recall status
Open, Classified
CDRH record ID
216634
FDA classified
2025-11-13
Recall event number
97814

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗