Falsely elevated B-hCG results can occur when running B12 and B-hCG assays on the same analyzer due to carryover. B-12 results are not impacted.
Source facts
Action
The firm initiated the recall via letter on December 14 and 16, 2004 directing consignees to perform the two tests on different machines or to contact the customer service representative in the event the consignee only has one machine.
Initiated
2001-12-14
Root cause
Other
Terminated
2005-11-02
Distribution
The B12 reagent was only distributed to foreign countries including Canada, Germany, Hong Kong, Japan, Dingapore, Australia, and New Zealand
Recall number
Z-0435-05
Recall status
Terminated
CDRH record ID
36550
FDA classified
2005-01-26
Recall event number
30764