Open original FDA record ↗Recall record
- FDA identifier
- Z-0435-2025
- Company in FDA record
- AVID Medical, Inc.
- FDA classification date
- Nov 14, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Source facts
- Action
- Avid Medical issued Urgent: Voluntary Medical Device Recall letter via email on 9/20/24. Letter states reason for recall, health risk and action to take: If you received any of the batches/lots identified in APPENDIX 1, please review your inventory, and discard the entire kit. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: If any of the affected kits identified in Appendix 1 were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. "Complete the attached Recall Response Form and return as soon as possible. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.
- Initiated
- 2024-09-20
- Root cause
- Process control
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0435-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210612
- FDA classified
- 2024-11-14
- Recall event number
- 95515