Open original FDA record ↗Recall record
- FDA identifier
- Z-0436-2011
- Company in FDA record
- Lemaitre Vascular, Inc.
- FDA classification date
- Nov 22, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Box label (Top Part). Lot and Expiration Date information were switched.
Source facts
- Action
- Lemaitre Vascular notified customers by an Urgent: VascuTape Radiopaque Tape Device Field Safety Notice letter on 10/22/10 via USPS with delivery confirmation. The Notice advised users of the incorrect lot and lot number on the outer box and also provided an image of the box and label for ease in identifying the product in their inventory. The letter states that the use and functionality of the device are not affected by the problem; however, if customers feel uncomfortable continuing to use the affected device, they can contact Customer Service for replacement. The firm recommended discarding the outer box of the affected product. The letter contains a form that is requested to be returned to LeMaitre Vascular as record of notification and reconciliation. Accounts can contact customer service at 800-628-9470.
- Initiated
- 2010-10-22
- Root cause
- Labeling Change Control
- Terminated
- 2011-09-15
- Distribution
- Nationwide Distribution -- AL, CA, CT, IL, ME, MI, MS, NC, NJ, NY, PA, PR, TX, and WA.
- Recall number
- Z-0436-2011
- Recall status
- Terminated
- CDRH record ID
- 95211
- FDA classified
- 2010-11-22
- Recall event number
- 57028