Open original FDA record ↗Recall record
- FDA identifier
- Z-0436-2016
- Company in FDA record
- Draeger Medical, Inc.
- FDA classification date
- Dec 23, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with the Infinity Workstation Critical Care (Evita Infinity V 500) were unexpectedly short due to the design of the charging algorithm in the current software. Devices used for patient transport will be a priority.
Source facts
- Action
- Draeger Medical, Inc. sent an URGENT MEDICAL DEVICE RECALL letter dated November 2015 to all affected consignees. The letter identified the product, problem, and actions to be taken by the customers. The Battery Charging Software will be updated and, if necessary, the batteries replaced in all affected devices Free of Charge.
- Initiated
- 2015-12-01
- Root cause
- Software design
- Terminated
- 2020-04-03
- Distribution
- US Nationwide-including DC and PR, and the states of AK; AR; AZ; CA; CO; CT; DE; FL; GA; HI; IA; ID; IL; IN; KS; LA; MA; MD; ME; MI; MN; MO; MS; MT; NC; ND; NE; NH; NJ; NM; NV; NY; OH; OK; OR; PA; SC; SD; TN; TX; UT; VA; VT; WA; WI; and WV.
- Recall number
- Z-0436-2016
- Recall status
- Terminated
- CDRH record ID
- 142153
- FDA classified
- 2015-12-23
- Recall event number
- 72754