Six of 370 stents released to market had a slightly below the required polymeric coating weight. All other specification were met.
Source facts
Action
A withdrawal letter is in the process of being hand-delivered to accounts that received this lot, by a Cordis sales representative. Any remaining units from this lot are being removed, and returned to Cordis. The returned units will be scrapped. An Acknowledgement Form is being signed indicating that the account was informed, and documenting whether there is any remaining product from this lot. The withdrawal action was initiated January 12, 2005.
Initiated
2005-01-12
Root cause
Other
Terminated
2005-02-09
Distribution
Distributed throughout the United States.
Recall number
Z-0437-05
Recall status
Terminated
CDRH record ID
36787
FDA classified
2005-01-26
Recall event number
30903