Open original FDA record ↗Recall record
- FDA identifier
- Z-0437-2013
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Nov 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Zimmer is initiating a lot specific recall of the NexGen Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset
Source facts
- Action
- Zimmer sent an Urgent Medical Device Recall letter dated September 21, 2012, to all affected customers. The letter identifiedthe product, the problem, and the action to be taken by the customer. Customers were instructed to locate and quarantine the affected product. Customers were to return the affected product along with the completed Inventory Return Certificate Form. If product was further distributed the Recall Letter should be forwarded to their customers. Product was to be returned to USA accounts - Zimmer Product Service Department, 1777 West Center St. Warsaw, IN 46580. If product has been implanted, affected surgeons will be notified. For questions regarding this recall call 574-372-4807.
- Initiated
- 2012-09-21
- Root cause
- Process change control
- Terminated
- 2014-04-17
- Distribution
- Worldwide Distribution - USA including AZ,CA, FL, IA, IL, IN, MO, NE, NJ, NY, OH, PA, WA. Internationally to Thailand, Canada, Germany, France, UK, Hungary, Netherlands, Russian Fed, and Sweden,
- Recall number
- Z-0437-2013
- Recall status
- Terminated
- CDRH record ID
- 114183
- FDA classified
- 2012-11-26
- Recall event number
- 63544