Open original FDA record ↗Recall record
- FDA identifier
- Z-0437-2014
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Dec 3, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A safety risk regarding a potentially existing dark current radiation phenomenon on the Linear Accelerator (LINAC) in combination with IMRT or mARC (rotational IMRT) treatments using unflat (flattening filter free) beams. In case there are many long periods of field shape changes in a treatment plan, the dose due to dark current radiation may sum up to a significant value.
Source facts
- Action
- Siemens sent an Urgent Medical Device Correction letter dated October 23, 2013 including the release of update instruction TH024/13/S to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer.
- Initiated
- 2013-10-23
- Root cause
- Device Design
- Terminated
- 2014-01-22
- Distribution
- Worldwide Distribution: USA (nationwide) including states of: FL, NC, NY, PA, TN, UT, and WI; and countries of: Angola, Brazil, Canada, P.R. China, Germany, India, Japan, Republic Korea, Poland, and Thailand.
- Recall number
- Z-0437-2014
- Recall status
- Terminated
- CDRH record ID
- 123180
- FDA classified
- 2013-12-03
- Recall event number
- 66724