Open original FDA record ↗Recall record
- FDA identifier
- Z-0438-2016
- Company in FDA record
- Medtronic MiniMed Inc.
- FDA classification date
- Dec 17, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.
Source facts
- Action
- Medtronic sent an Urgent Medical Device Recall Notification dated September 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letters stated that the pump drive motors may experience a malfunction which would result in a pump error message alarm notifying the patient that insulin is no longer being delivered. The letters also inform the customers of what actions are to be taken. Customers with questions are instructed to contact Medtronic MiniMed's customer support team. For further questions please call (818) 576-4700.
- Initiated
- 2015-09-25
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-02-05
- Distribution
- Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom.
- Recall number
- Z-0438-2016
- Recall status
- Terminated
- CDRH record ID
- 140715
- FDA classified
- 2015-12-17
- Recall event number
- 72352