Skip to content

Public FDA record

Z-0438-2024: CardioQuip Modular Cooler-Heater, Model:MCH-1000(m), Power: 115V-60hz,16A

Recall record Z-0438-2024 for CardioQuip, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0438-2024
Company in FDA record
CardioQuip, LLC
FDA classification date
Dec 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components.

Source facts

Action
On 10/19/23, CardioQuip issued a "Urgent: Medical Device Correction" notification to all affected consignees. This notification was an update to the firm's May 22, 2023 communication. CardioQuip ask consignees to take the following actions: 1. Prime the MCH-1000(i) and MCH-1000(m) prior to each use. 2.Never operate the device dry (see communication for further details). 3. Design Changes to be implemented (see communication for further details). A. Relocation of the water Level Sensor (see communication for further details). B. Software Update (see communication for further details). 4. Please fill out and submit the receipt of understanding of the contents of this letter to CardioQuip within 5 business days or receipt. 5. Any adverse events experienced relating to CardioQuip devices should be reported to CardioQuip or through the FDA s MedWatch Program: Phone: +1 (800) FDA-1088 Web: www.fda.gov/medwatch/report.htm
Initiated
2023-10-19
Root cause
Device Design
Distribution
US Distribution: TX, IN, IL, WI, FL, NJ, PA, KY, VA, WV, CT, CA, WA, AL, IA
Recall number
Z-0438-2024
Recall status
Open, Classified
CDRH record ID
204054
FDA classified
2023-12-01
Recall event number
93347

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗