The roller pump will pause in response to a pressure alarm condition, as designed, but then may not resume operation as expected after the alarm condition has been cleared.
Source facts
Action
Consignees were notified of the problem via letter dated 11/10/06 and informed that they would be contacted to schedule a service call to install software upgrades. Installation of software upgrades began at customer locations on 10/16/06.
Initiated
2006-10-16
Root cause
Other
Terminated
2009-07-07
Distribution
Worldwide distribution ---- including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Recall number
Z-0439-2007
Recall status
Terminated
CDRH record ID
48742
FDA classified
2007-02-14
Recall event number
36494