Open original FDA record ↗Recall record
- FDA identifier
- Z-0439-2013
- Company in FDA record
- Cooper Surgical, Inc.
- FDA classification date
- Nov 26, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle
Source facts
- Action
- CooperSurgical issued notification on October 31, 2012, to all affected customers Fed X . The notification identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return units for replacement. Customers with questions should contact Customer Serivce at 1-203-601-9818 or 800-243-2974. Customers were asked to fill out the attached Replacement of Recalled Product Form and fax it to 1-800-262-0105. For questions regarding this recall call 203-601-9818.
- Initiated
- 2012-10-31
- Root cause
- Device Design
- Terminated
- 2014-03-04
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Belgium, Canada, Czech Republic, Germany, Italy, Spain, France, United Kingdom, Japan, New Zealand, South Korea, Saudi Arabia, Sweden, Singapore, Slovenia, and Taiwan.
- Recall number
- Z-0439-2013
- Recall status
- Terminated
- CDRH record ID
- 114522
- FDA classified
- 2012-11-26
- Recall event number
- 63681