Z-0439-2021: Liberty Select Cycler Software Version 2.9.0 Model Number 180343 and RTLR180343

Recall record Z-0439-2021 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0439-2021
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA record creation date
Nov 7, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device may detect an incorrect Heater Bag volume which may lead to a 'Supply Bag Line Blocked' alarm during treatment. This alarm may cause the Heater Bag to contain more fluid than expected and the cycler does not recognize the extra fluid is present and it will not make it available for treatment. Not having enough solution to complete treatment may lead to inadequate dialysis which could result in the accumulation of toxins and fluid. Symptoms may include shortness of breath, increased weight gain, and increased blood pressure.

Source facts

Action
On September 16, 2020, Urgent Medical Device Correction notices and response forms were issued to home therapy nurse managers and patients. Patients using a Liberty Select cycler with software version 2.9.0 are instructed to disable the Flow Alert option. Patients will no longer receive on an-screen Flow Alert message or an audible Flow Alert series of beeps. Instructions for disabling the Flow Alert alarm are included in the notification letter. Patients should continue to use their Liberty Select Cycler for treatment with the Flow Alert option set to "No". Fresenius Medical Care LLC is not requesting the return of the Liberty Select cyclers software version 2.9.0. Replacing the Liberty Select cycler will not resolve this issue. Please complete and return the enclosed Reply Form, indicating receipt and understanding of the communication and actions to take. For additional information, please call 1-855-616-2309 Monday - Friday 8:30 am-5 pm, EST or submit an online request any time at www.fmcna-medinfo.com.
Initiated
2020-09-12
Root cause
Device Design
Distribution
Distributed nationwide throughout the U.S and Puerto Rico. 9,907 devices currently in the field. 361 devices have been deactivated. 4,870 devices are ready to be shipped to customers along with the Urgent Medical Device Correction notice and response form.
Recall number
Z-0439-2021
Recall status
Completed
CDRH record ID
183952
FDA record created
2020-11-07
Recall event number
86537

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗