Reason stated in the source

A capacitor of the wrong value was used. The capacitor used was 0.11 microfarad instead of 0.22 microfarad called for in the product specifications.

Source facts

Action
The recalling firm notified hospitals and distributors of the product by letter, dated 12/10/04.
Initiated
2004-12-10
Root cause
Other
Terminated
2008-10-29
Distribution
The product was distributed to consignees located in MI, TN, IL, NC, GA, IL, CA, FL, IN, WI, TX, AR, KS, IA, OK, NY, PA, MS, KY, ME, MO, MA, VA, NJ, LA, MN, SD, UT, CO, AL, and NH. There are no foreign consignees.
Recall number
Z-0440-05
Recall status
Terminated
CDRH record ID
36451
FDA classified
2005-01-27
Recall event number
30718

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗