A capacitor of the wrong value was used. The capacitor used was 0.11 microfarad instead of 0.22 microfarad called for in the product specifications.
Source facts
Action
The recalling firm notified hospitals and distributors of the product by letter, dated 12/10/04.
Initiated
2004-12-10
Root cause
Other
Terminated
2008-10-29
Distribution
The product was distributed to consignees located in MI, TN, IL, NC, GA, IL, CA, FL, IN, WI, TX, AR, KS, IA, OK, NY, PA, MS, KY, ME, MO, MA, VA, NJ, LA, MN, SD, UT, CO, AL, and NH. There are no foreign consignees.
Recall number
Z-0440-05
Recall status
Terminated
CDRH record ID
36451
FDA classified
2005-01-27
Recall event number
30718