Open original FDA record ↗Recall record
- FDA identifier
- Z-0440-2014
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Dec 4, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip of the dilator has separated fro the body of the dilator.
Source facts
- Action
- Urgent recall notification letters were sent via Federal Express Priority mail on November 5, 2013. The notification instructs the account to check their inventory, remove any affected product from their inventory, and return it to Boston Scientific. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product in their inventory. Follow-up communications will be made to the accounts that do not respond to this inquiry.
- Initiated
- 2013-11-05
- Root cause
- Component design/selection
- Terminated
- 2014-12-08
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Ireland, Denmark, Italy, France, Canada, Great Britain, Switzerland, South Africa, Germany, and the Netherlands.
- Recall number
- Z-0440-2014
- Recall status
- Terminated
- CDRH record ID
- 123623
- FDA classified
- 2013-12-04
- Recall event number
- 66828