Pump not detected as connected to controller due to software issue.
Source facts
Action
An Urgent Medical Device Correction was issued vail mail October 24. 2023. Customers are advised to examine inventory for product running AIC software version: V8.4 or V8.4.1. Do not to use the product unless software is updated. Contact adminsfs@abiomed.com to schedule the software update and return the business reply form. Forward this information to relevant personnel and any facility product may have been transferred to.
Initiated
2023-10-10
Root cause
Software design
Terminated
2025-03-06
Distribution
US: AZ, CA, FL, NY & VA.
Recall number
Z-0440-2024
Recall status
Terminated
CDRH record ID
204275
FDA classified
2023-12-01
Recall event number
93406