Z-0441-2021: Strep B Carrot Broth One-Step, Ref. Z40, UDI: 00819398026499

Recall record Z-0441-2021 for Hardy Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0441-2021
Company in FDA record
Hardy Diagnostics
FDA record creation date
Nov 10, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Streptococcus culture medium contains an incorrect expiration date. The culture medium outer box labeling contains an expiration date of 2021-11-10 instead of the correct expiration date of 2020-11-10

Source facts

Action
Hardy Diagnostic sent a "Urgent: Recall Notification" on 9/30/2020 and sent a follow-up letter 10/06/2020 containing additional information. In addition to notifying consignees about the recall, the firm asked consignees to take the following actions: 1. Please respond to the notification within 10 business days of receipt of the letter, either through mail or email. 2. To respond by email, please scan and send the completed inventory reconciliation form (FORM-2023C) to TechService@HardyDiagnostics.com. After a response is received, new labels with the correct expiration date will be sent to you for product that has not yet been used. 3. Upon expiration of the product, on November 10, 2020, discard any remaining stock. This product should be destroyed according to local regulation as non-hazardous waste if un-inoculated or as medical waste if inoculated. 4. Please retain this letter or post the letter where the product is stored to ensure the users are aware of this defect. Additionally, please take the following actions: - Notify any and all employees who may use the product. - If the product was further distributed, notify the consignees the product was distributed to. 5. Any Adverse reactions or quality problems experienced with the use of this product may be reported to the FDAs MedWatch Adverse Event Reporting program either online (www.fda.gov/medwatch/report.htm), by regular mail or by fax. 6. Please feel free to contact Technical Services if you have additional questions about this notification through email (listed above) or through phone call at (800) 266-2222, option 2, Monday through Friday, 8:00 AM to 5:00 PM, Pacific Time.
Initiated
2020-10-06
Root cause
Employee error
Terminated
2021-05-28
Distribution
US: VA, AR, WI, ID, NY, AL, CA, MT, TX, MN, NH, WA, NC, NY, AZ, AK, MO, FL, MI, WY, ID, CO, AL, MT, WY, Puerto Rico, and Guam
Recall number
Z-0441-2021
Recall status
Terminated
CDRH record ID
184105
FDA record created
2020-11-10
Recall event number
86586

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗