- Recall classification
- Class II
Reason stated in the source
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Source facts
- Action
- Customers were notified via FedEx and/or email on approximately 08/07/2017. Instructions include ensure appropriate personnel are notified of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide the names of hospitals or customers that would have received this product if further distributed.
- Initiated
- 2017-08-07
- Root cause
- Process change control
- Terminated
- 2019-03-29
- Distribution
- US nationwide distribution. Also Portugal and the Netherlands.
- Recall number
- Z-0442-2018
- Recall status
- Terminated
- CDRH record ID
- 160340
- FDA record created
- 2018-01-29
- Recall event number
- 78763