Z-0442-2022: Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA T

Recall record Z-0442-2022 for Karl Storz Endoscopy. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0442-2022
Company in FDA record
Karl Storz Endoscopy
FDA record creation date
Dec 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality.

Source facts

Action
On 10/15/21 correction notices were mailed to customers. Customers were asked to do the following: 1) Ensure all personnel in your organization review this letter and the current instructions for use. If you have previously sold or transferred affected devices to a third party, please forward this notice to all such third parties. 2) Discard any prior versions of the instructions for use. 3) Complete and return the acknowledgment form. Customers with medical or safety-related inquiries can contact Medical Affairs at 1-800-649-8852 or medicalaffairs@karlstorz.com. Customers with questions about the correction communication are encouraged to call 1-866-913-0667 or email karlstorz7824@stericycle.com.
Initiated
2021-10-15
Root cause
Device Design
Distribution
US: AK, AL, AP, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: Canada, Venezuela, Barbados, GUAM, Puerto Rico, Mexico, Northern Mariana Islands, Trinidad & Tobago, Germany, Cayman Islands, Philippines, Japan, Dominican Republic, Panama, Bahamas, Ecuador, Suriname, American Virgin Islands, British Virgin Islands, Belize, Curacao, The Bahamas, Brazil, South Korea, Bermuda
Recall number
Z-0442-2022
Recall status
Open, Classified
CDRH record ID
190238
FDA record created
2021-12-29
Recall event number
88998

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗