Open original FDA record ↗Recall record
- FDA identifier
- Z-0443-06
- Company in FDA record
- Becton Dickinson & Co.
- FDA classification date
- Jan 25, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A operator's manual for a diagnostic test kit is incorrect and may cause inaccurate results for bacterial identification and antibiotic susceptibility in patient samples.
Source facts
- Action
- The recalling firm notified direct accounts by courier and international distributors by e-mail on 01/17/06. The notification advised that the the troubleshooting section in the firm''s nephelometer user''s guide, vers. 2004/12, is incorrect in describing the 'CAL?' message as indicating a default calibration is in use. The recalling firm provided a cautionary sticker in the bulletin and described recommended placement on the device by pictorial. The sticker instructs users not to use the nephelometer if 'CAL?' message appears on the units'' LCD screen.
- Initiated
- 2006-01-17
- Root cause
- Other
- Terminated
- 2006-05-05
- Distribution
- The devices were distributed to direct accounts nationwide and affiliated distributor facilities worldwide.
- Recall number
- Z-0443-06
- Recall status
- Terminated
- CDRH record ID
- 43774
- FDA classified
- 2006-01-25
- Recall event number
- 34396