Skip to content

Public FDA record

Z-0444-05: Internal Battery for Puritan Bennett 700 Series Ventilator.

Recall record Z-0444-05 for Nellcor Puritan Bennett Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0444-05
Company in FDA record
Nellcor Puritan Bennett Inc
FDA classification date
Jan 28, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Wires on battery may loosen causing disconnection.

Source facts

Action
Letters will be Fedexed to customers which includes a simple test instruction that users can use to test the performance of the back up alarm.
Initiated
2004-12-16
Root cause
Other
Terminated
2008-01-23
Distribution
Nationwide, Europe, Japan, Latin America and other portions of Asia.
Recall number
Z-0444-05
Recall status
Terminated
CDRH record ID
36592
FDA classified
2005-01-28
Recall event number
30779

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗