The instructions for use contain pre-vacuum sterilization instructions that may lead to compromised sterility at the lid-kit interface location.
Source facts
Action
Zimmer sent out "Urgent Device Correction" letters dated 5/18/2010. The consignees were provided with new instructions for use. The letter informed users that the current pre-vacuum cycle of 3 to 18 minutes is not enough time to assure sterility between the lid-kit interface and could lead to cross contamination of multiple patients. The users were informed that a pre-vacuum time of 30 minutes was required to assure sterility.
Initiated
2010-05-18
Root cause
Labeling design
Terminated
2011-05-18
Distribution
Distribution Nationwide, to France, and Spain.
Recall number
Z-0444-2011
Recall status
Terminated
CDRH record ID
92220
FDA classified
2010-11-23
Recall event number
55889