Z-0444-2022: Catalys Precision Laser System, Models: Catalys-U (United States), Catalys-I (International), Catalys-C (

Recall record Z-0444-2022 for AMO Manufacturing USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0444-2022
Company in FDA record
AMO Manufacturing USA, LLC
FDA record creation date
Dec 29, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.

Source facts

Action
On 09/02/21, Microsoft "PrintNightmare" Vulnerability notices were sent to customers informing them of the upcoming software patches to remediate the vulnerability. This notice was posted on the recalling firm's website in the Security Advisories section. On 10/21/21 correction notices started being sent to customers. Customers were informed that Technical Service Representatives would contact then to schedule installation of the Windows OS patch. Customers were asked to do/ensure the following: 1) Complete the customer reply form, 2) To safeguard patients form potential harm and to inform them what to do in the event of product malfunction, customers were directed to specific parts of the Instructions for Use. 3) Confirm all measurements against baseline pre-operative manifest refraction and do not proceed when outside of recommended limits of difference, for sphere, cylinder and axis, as specified in the operators manual and training material. 4) Maintain manual records of patient measurements. 5) Ensure you comply with obligations under applicable privacy laws relating to putting in place safeguards in connection with personal information for patients. Customers with additional questions were encouraged to call Technical Support at 1 (800)- 511-0911.
Initiated
2021-10-21
Root cause
Software design
Distribution
US: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM, CA. OUS: Canada, Mexico, Russian Federation, Turkey, Brazil, Egypt, Iraq, Saudi Arabia, Jordan, Norway, Colombia, Germany, Malaysia, India, Oman, Azerbaijan, Canada, Slovakia, Lebanon, Spain, Ireland, Italy, South Korea, Bangladesh, Chile, Taiwan, Tunisia, Kazakhstan, Libya, Argentina, Singapore, Austria, Australia, Vietnam, Greece, Poland, Indonesia, Yemen, Great Britain, South Africa, Croatia, France, Czech Republic, Algeria, Thailand, Bulgaria, United Arab Emirates, Belgium, Japan, China, Nepal, Israel, Switzerland, Dominican Republic, Hong Kong, Portugal, Ecuador, Kuwait, Netherlands, Slovenia, Philippines
Recall number
Z-0444-2022
Recall status
Open, Classified
CDRH record ID
190195
FDA record created
2021-12-29
Recall event number
88986

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗