Skip to content

Public FDA record

Z-0445-2022: Invacare TDX SP2 with LiNX Gyro Power Wheelchair (with LiNX Gyro Module Model Numbers DLX-GRY100-A and DL

Recall record Z-0445-2022 for Invacare Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0445-2022
Company in FDA record
Invacare Corporation
FDA record creation date
Jan 3, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Power Wheelchairs with a LiNX Gyro component running firmware version 6.1.2 can experience a more aggressive deceleration rate than the programmed rate due to the system following the incorrect deceleration profile, resulting in potential injury

Source facts

Action
Invacare issued Urgent Medical Device Letter issued on 11/12/21 to consignees who received devices with LiNX Gyro modules containing firmware version 6.1.2. This letter also contains instructions for consignees who upgraded devices in the field to version 6.1.2. A Consumer Letter will be provided on the recall website for Providers to send to end users who are impacted. Letter states reason for recall, health risk and action to take: Visit www. Invacare.com/RC916 within five (5) business days of receiving this notice. Enter the requested identification information to access the list of power wheelchair serial numbers and/or LiNX Gyro Modules sold to you and instructions to conduct this field correction. You will be prompted to provide responses to information requested regarding this field correction and affected units. " The account number required to access your list of impacted devices appears directly at the end of your company name in the mailing address. For example, ABC COMPANY-XXXXX. " The mailing postal code will also be required to log into your account. 2. Stop using LiNX Gyro Module firmware version 6.1.2 on all power wheelchair programming tools and stored electronic media. 3. Immediately update any LiNX Access iOS and LiNX Access PC tools containing LiNX Gyro Module firmware version 6.1.2 to the latest LiNX Access iOS or LiNX Access PC tool version. 4. Locate the power wheelchairs with affected LiNX Gyro Modules per the date ranges above. 5. For power wheelchairs built prior to the date ranges above, review your customer records to identify whether you have installed LiNX firmware version 6.1.2 on any LiNX Gyro-equipped power wheelchairs since the January 14, 2021 release date of firmware version 6.1.2. Note: If you do not have record of the firmware version installed on the LiNX Gyro Modules you have purchased from Invacare, you must connect to the LiNX system on the wheelchair using the LiNX Access tool to determine the firmware version of the Gyro
Initiated
2021-11-12
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Thailand, New Zealand.
Recall number
Z-0445-2022
Recall status
Open, Classified
CDRH record ID
190405
FDA record created
2022-01-03
Recall event number
89062

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗