Open original FDA record ↗Recall record
- FDA identifier
- Z-0446-2007
- Company in FDA record
- Respironics, Inc.
- FDA classification date
- Jan 10, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
AC inlet connector failure that can result in thermal damage to the humidifier enclosure and/or ignition of materials external to the device
Source facts
- Action
- The recalling firm issued a recall letter dated 10/25/06 to their domestic and Canadian dealers informing them of the problem and the need to contact the end users. These dealers were given two options: 1) contact the end user by forwarding a letter the recalling firm provided or 2) contacting the third party company with their distribution list. The recalling firm issued a recall letter dated 10/25/06 regarding the service repair kit and the need to contact end users that had the device repaired using this kit. The recalling firm issued a recall letter date 10/25/06 to the international customers informing them of the problem, contact the end users, and the need to destroy the recalled part.
- Initiated
- 2006-10-25
- Root cause
- Other
- Terminated
- 2009-12-15
- Distribution
- Worldwide Distribution ---- including USA and countries of Denmark, Austria, France, Canada, Netherlands, Japan, Hong Kong, Austrailia, Great Britain, Spain, Belgium, Germany, Poland, Ireland, Portugal, Saudia Arabia, Korea, Mexico, Finland, Taiwan, Philippines, Iceland, Brazil, Kuwait, New Zealand, Chile, India, Egypt, Norway, Turkey, Jordan, Thailand, Hungary, Russia, Peru, South Africia, Israel, Czech, Pakistan, Lebanon, Venezuela, Sweden, United Arab Emirates, Qatar, Slovakia, Bulgaria, Bermuda, and Colombia.
- Recall number
- Z-0446-2007
- Recall status
- Terminated
- CDRH record ID
- 49054
- FDA classified
- 2007-01-10
- Recall event number
- 36649