Open original FDA record ↗Recall record
- FDA identifier
- Z-0447-2008
- Company in FDA record
- Dale Medical Products, Inc.
- FDA classification date
- Jan 11, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Revised Labeling : Instructions For use: "Do not trim fastener hook tabs. " Trimming can cause tabs to disengage at the trach plate
Source facts
- Action
- Dale Medical notified Distributors by letter dated 9/27/07, identified as an Urgent Medical Device Correction Letter. The letter requested distributors to notify all their customers by letter. The Correction Letter reminded users to take notice of new labeling " Do not trim fasterner hook tabs"
- Initiated
- 2007-09-27
- Root cause
- Labeling design
- Terminated
- 2012-05-09
- Distribution
- Worldwide-USA and countries of Canada, Austria, Belgium, Denmark, Finland, France, Germany, Greece, Iceland, Italy, The Netherlands, Norway, The UK, Spain, Switzerland, Australia, Hong Kong,India, Japan, Korea, Malaysia, New Zealand, The Philippines,Singapore, Taiwan, ROC, Brazil, Bermuda, Chile, Costa Rica,Ecuador, Mexico, Panama, Venezuela, Thailand, Saudi Arabia, and The United Arab Emirates.
- Recall number
- Z-0447-2008
- Recall status
- Terminated
- CDRH record ID
- 65505
- FDA classified
- 2008-01-11
- Recall event number
- 45460