The packages for Modular Shoulder Body Assembly Left (SH-1540L-S) and Modular Shoulder Body Assembly Right (SH-1540R-S) may contain assemblies for the opposite side as indicated.
Source facts
Action
The firm issued a Urgent Notice: Device Recall on 5/13/08. The firm notified consignees of the issue and asked them to return the product to Acumed with the enclosed product recall effectiveness form for customers. For further information contact Acumed customer service at 1-888-627-9957
Initiated
2008-05-13
Root cause
Labeling mix-ups
Terminated
2010-05-13
Distribution
World-wide distribution including states of GA, MD, CA, MO, and AL, countries of United Kingdom, South Korea, South Africa, and Spain.
Recall number
Z-0447-2009
Recall status
Terminated
CDRH record ID
71485
FDA classified
2008-12-11
Recall event number
48470