Z-0447-2021: Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended t

Recall record Z-0447-2021 for Shanghai United Imaging Healthcare Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0447-2021
Company in FDA record
Shanghai United Imaging Healthcare Co., Ltd.
FDA record creation date
Nov 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Source facts

Action
The firm sent notification letters via email to distributorsThe firm's consignee hand delivered a written Field Safety Notification to end-users on June 11, 2020 to inform them that the delay time displayed on the DDP can be incorrect for a timed scan such as a multi-phase contrast scan. The DDP is located on the gantry in the scan room. If the operator enters the scan room to perform any multi-phase contrast scan and refers to the incorrect delay time displayed on the DDP, he or she may not be able to leave the scan room before the scan starts, and as a result may receive unnecessary radiation dose. The letter states the following: The purpose of this letter is to advise you that United Imaging is voluntarily correcting some of the uCT 530, uCT 550, uCT 760 The uCT 530 and uMI 550 devices. Reason for the Voluntary Correction: We are aware of one complaint associated with the device. We have not received any report that have resulted in injury. Safety Issue: A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. Risk to Health of Safety Issue: The delay time displayed on the DDP can be incorrect for a timed scan such as a multi-phase contrast scan. The DDP is located on the gantry in the scan room. If the operator enters the scan room to perform any multi-phase contrast scan and refers to the incorrect delay time displayed on the DDP, he or she may not be able to leave the scan room before the scan starts, and as a result may receive unnecessary radiation dose. Actions to be taken by the Customer I User: 1. During a multi-phase contrast scan, do not perform any operations on the Preview image of any previous phases. 2. If possible, avoid entering the scan room when performing multi-phase contrast scans. 3. If it is necessary to enter the scan room when performing a multi-phase contrast scan, ignore the potentially incorrect delay time displayed on the
Initiated
2020-08-05
Root cause
Software design
Distribution
US Nationwide distribution including in the state of TX.
Recall number
Z-0447-2021
Recall status
Open, Classified
CDRH record ID
183279
FDA record created
2020-11-12
Recall event number
86269

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗