Skip to content

Public FDA record

Z-0447-2025: Halyard ANGIOGRAPHY DRAPE PACK - Medical convenience kits Model Number: VMCC021-06

Recall record Z-0447-2025 for AVID Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0447-2025
Company in FDA record
AVID Medical, Inc.
FDA classification date
Nov 14, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Source facts

Action
Avid Medical issued Urgent: Voluntary Medical Device Recall letter via email on 9/20/24. Letter states reason for recall, health risk and action to take: If you received any of the batches/lots identified in APPENDIX 1, please review your inventory, and discard the entire kit. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: If any of the affected kits identified in Appendix 1 were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. "Complete the attached Recall Response Form and return as soon as possible. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com.
Initiated
2024-09-20
Root cause
Process control
Distribution
US Nationwide distribution.
Recall number
Z-0447-2025
Recall status
Open, Classified
CDRH record ID
210628
FDA classified
2024-11-14
Recall event number
95515

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗