Open original FDA record ↗Recall record
- FDA identifier
- Z-0448-2007
- Company in FDA record
- Drew Scientific Group PLC
- FDA classification date
- Jan 9, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Flags and alerts on the user interface will not automatically transfering to the LIS.
Source facts
- Action
- Firm sent 'Product Notification #22' to all customers on 08/24/06. Users advised of potential risk to patient health and provided with instructions on how to eliminate risk by disabling the auto transfer feature and manually transfering data until a software patch is made available
- Initiated
- 2006-08-24
- Root cause
- Other
- Terminated
- 2008-03-10
- Distribution
- Worldwide distribution --- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India, Iran, Korea, Mexico, Morocco, Phillipines, Portugal, Russia, Serbia and Montenegro, Spain, Syria, Thailand, Turkey, United Arab Emerates (Dubai), United Kingdom and Vietnam, West Indies and Yemen.-
- Recall number
- Z-0448-2007
- Recall status
- Terminated
- CDRH record ID
- 49192
- FDA classified
- 2007-01-09
- Recall event number
- 36706