A high homocysteine level can be reported as falsely low.
Source facts
Action
On 01/18/07 an "URGENT: PRODUCT CORRECTIVE ACTION" letter was sent to the consignees. The letter states the circumstances in which a high homocysteine rate can result in a false low homocysteine rate, a work around for the problem which a flag must be programmed into the user defined homocysteine test parameters and instructions to advise their customers about the situation was addressed in the letter so the consignees can take the recommended corrective action. A confirmation sheet was included with the letter to be signed by a representative of each distributor and returned to Catch Inc. Contact Catch Inc. at 1.425.402.8960 for assistance.
Initiated
2007-01-18
Root cause
Device Design
Terminated
2010-09-28
Distribution
CA and PA.
Recall number
Z-0448-2009
Recall status
Terminated
CDRH record ID
72016
FDA classified
2009-01-06
Recall event number
48734