- FDA identifier
- Z-0448-2015
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA classification date
- Nov 25, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
Source facts
- Action
- Urgent Medical Device Recall letters were sent to affected sites on November 3, 2014. Letters inform customers of the reason for recall, risk to health and the affected instrument. A table of all of the affected IS4000 instruments affected is attached to the letter indicating the part number and product name for each. Letters also inform customers that all instruments shipped since July 2014 are manufactured with a new material and not subject to recall. Customers are to take the following actions: 1. Ensure that all affected personnel are aware of the information. 2. Identify affected product in their possession 3. Contact ISI rep for replacement. 4. Complete and return the attached Acknowledgement From. Until replacement instruments are received, the old instruments may be used provided they do not show signs of degradation and that the reprocessing instructions are followed. Questions to 800-876-1310 Option 3 ( 6 am to 5 pm PST) or ics@intusurg.com.
- Initiated
- 2014-11-03
- Root cause
- Labeling design
- Terminated
- 2015-02-03
- Distribution
- US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.
- Recall number
- Z-0448-2015
- Recall status
- Terminated
- CDRH record ID
- 131248
- FDA classified
- 2014-11-25
- Recall event number
- 69638