Z-0449-2024: Monarch Platform REF MON-000005-01R, 100-240V- 50/60Hz 1440w

Recall record Z-0449-2024 for AURIS HEALTH INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0449-2024
Company in FDA record
AURIS HEALTH INC
FDA classification date
Dec 1, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is the potential that software issues may result in flipped image of the virtual Bronchoscope view.

Source facts

Action
On October 19, 2023, Auris Health, Inc. issued an "URGENT: MEDICAL DEVICE CORRECTION" notification to affected consignees. Auris asked consignees to take the following actions: Should an image flip be detected in the Virtual Bronchoscope view while navigating toward the target, please perform the following steps to correct the image orientation: 1. Retract the MONARCH Bronchoscope outside the patient 2. Undock the MONARCH Bronchoscope from the arms 3. Clear any fault pop-up messages 4. Re-dock the MONARCH Bronchoscope to the arms 5. Re-initialize Navigation and proceed with the case 6. Please review this letter carefully and share it with anyone in your facility that needs to be informed. 7. Please complete all fields of the attached Acknowledgement Form confirming receipt of this notification and return the form via email to MonarchSupport@its.jnj.com. If you have any questions regarding this letter or need to report any customer complaints, please contact the MONARCH Support team at +1 (800) 434-0032 or via email at MonarchSupport@its.jnj.com.
Initiated
2023-10-18
Root cause
Software design
Distribution
US Distribution: AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WI & WV.
Recall number
Z-0449-2024
Recall status
Open, Classified
CDRH record ID
203933
FDA classified
2023-12-01
Recall event number
93320

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗