Z-0450-2021: VITROS Immunodiagnostic Products-SARS-CoV-2 IgG Calibrator - Product Usage: In-Vitro Diagnostic (EUA) an

Recall record Z-0450-2021 for Ortho Clinical Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0450-2021
Company in FDA record
Ortho Clinical Diagnostics Inc
FDA record creation date
Nov 13, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Elevated sample results signal/cutoff (s/c) values when using eight (8) lots of VITROS immunodiagnostic Products Anti-SARS-CoV-2 IgG Calibrator and decreased signal in one (1) lot of VITROS Immunodiagnostic Products Anti-SARS-CoV-2 IgG Control

Source facts

Action
Ortho Clinical Diagnostics issued Urgent Product Correction Notice on 30 Sept 2020 via(CL2020-246) of the issue and instructed as follows: If they are unable to calibrate (now or in the future) using an affected lot of VITROS Anti-SARS-CoV-2 IgG Calibrator Ortho will replace their remaining inventory as indicated on their Confirmation of Receipt form. NOTE: If they have a successful calibration, and quality control results are acceptable, the laboratory may continue using the affected lots for qualitative detection of IgG antibodies to SARS-CoV-2. If they are using the affected product to qualify units as high titer for CPT, consult with your medical director to determine if they should continue testing while waiting for their replacement product. Ortho will replace or credit their remaining inventory as indicated on their Confirmation of Receipt form. Post this notification by each system that processes the affected lots of VITROS Anti-SARS-CoV-2 IgG Calibrator or Control. Complete the enclosed Confirmation of Receipt form no later than October 9, 2020. Indicate the quantity of their remaining inventory to be replaced or credited. Please forward this notification if the product was distributed outside of their facility
Initiated
2020-09-30
Root cause
Nonconforming Material/Component
Terminated
2023-12-07
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Brazil, Chile, China, Colombia, India, Japan, Singapore 609917, United Kingdom, Italy.
Recall number
Z-0450-2021
Recall status
Terminated
CDRH record ID
184069
FDA record created
2020-11-13
Recall event number
86573

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗