System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts.
Source facts
Action
Firm sent 'Technical Bulletin #39' to all customers on 09/20/06. Users advised of potential for inaccurate readings and provided with instructions on how to eliminate problem.
Initiated
2006-09-20
Root cause
Other
Terminated
2008-03-10
Distribution
Worldwide distribution ---- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India, Iran, Korea, Mexico, Morocco, Phillipines, Portugal, Russia, Serbia and Montenegro, Spain, Syria, Thailand, Turkey, United Arab Emerates (Dubai), United Kingdom and Vietnam, West Indies and Yemen.
Recall number
Z-0451-2007
Recall status
Terminated
CDRH record ID
49231
FDA classified
2007-01-05
Recall event number
36728