Open original FDA record ↗Recall record
- FDA identifier
- Z-0451-2008
- Company in FDA record
- Philips Medical Systems
- FDA classification date
- Dec 11, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Alarm failure : If changes are made to the Caregroup Editor configuration, Care Group alarm status, including alarm text and prompt tones, may fail to be displayed or annunciated at bedside monitors within the Care Group
Source facts
- Action
- Philips Medical notified customers by letter dated November 2007 titled : Urgent Medical Device Correction Notice. Customers will be instructed to follow the Procedure to Mitigate Risk section * of the notice while they await the correction of their devices with a mandatory software upgrade to be performed by service representatives. * (no changes are to be made to the Caregroup Editor configuration on any system. If changes have already been made to the configuration prior to receipt of this notification, turn each bedside monitor "off " and then "on" in order to reset the system-Customer Care Contact telephone 1-800-722-9377)
- Initiated
- 2007-11-13
- Root cause
- Software change control
- Terminated
- 2016-08-15
- Distribution
- Worldwide Distribution -USA including states of CA, MA, TX, IN, MI, FL, MN, NC, MS, HI, RI, NH, and PA, and countries of Canada, Australia, Belgium, France, Germany, Ireland, Italy, Japan, Switzerland, United Arab Emirates, and The United Kingdom.
- Recall number
- Z-0451-2008
- Recall status
- Terminated
- CDRH record ID
- 66252
- FDA classified
- 2007-12-11
- Recall event number
- 45843