Skip to content

Public FDA record

Z-0451-2014: Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothromb

Recall record Z-0451-2014 for Abbott Point Of Care Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0451-2014
Company in FDA record
Abbott Point Of Care Inc.
FDA classification date
Dec 4, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.

Source facts

Action
Abbott Point of Care Inc. (APOC) issued an Urgent Recall Notice on October 28, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of cartridges from PT/INR lots in this letter. All unused cartridges would be credited and are to be returned as instructed on the enclosed Business Reply Card. If customers provided cartridges from the lot number listed above to another institution they were instructed to provide a copy of this letter to them. Customers with questions were instructerd to contact Abbott Point of Care Technical at 1-800-366-8020, Option 1. For questions regarding this recall call 800-366-8020.
Initiated
2013-10-28
Root cause
Nonconforming Material/Component
Terminated
2014-05-29
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Brazil, Chile, China, Finland, France, Germany, Greece, Hong Kong, Ireland, Italy, South Korea, Kuwait, Mexico, Netherland, New Zealand, Nigeria, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, south Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE, and United Kingdom,
Recall number
Z-0451-2014
Recall status
Terminated
CDRH record ID
123293
FDA classified
2013-12-04
Recall event number
66757

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗