Open original FDA record ↗Recall record
- FDA identifier
- Z-0452-2016
- Company in FDA record
- Zimmer, Inc.
- FDA classification date
- Dec 21, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm identified a malfunction which made it possible for a handpiece to start by itself when the power source is connected.
Source facts
- Action
- Zimmer sent an Urgent Medical Device Correction letter dated November 9, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Review this notification. 2. You will receive copies of notification sent directly to hospital risk managers and surgeons in your territory. As necessary, review and facilitate understanding of this notification by those entities. 3. Further identify hospital risk managers and/or surgeons in your territory who should also receive notification of this update, and supply the information to the entities you have identified. 4. If after reviewing this notification you have further questions or concerns, please call 330-364-0989 between the hours of 8 a.m. and 5 p.m. EST, Monday through Friday. Zimmer instructed their direct customers that they will upgrade the device with the new electronic during their annual preventative maintenance activities. Until the new upgrade is installed, the firm instructed their customers to follow the Instructions for Use.
- Initiated
- 2015-11-10
- Root cause
- Device Design
- Terminated
- 2016-02-19
- Distribution
- US Distribution to the states of : AR, CA, CO, CT, FL, GA, HI, IL, KS, KY, LA, ME, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, PA, RI, SC, TX, UT, VA, WA, WI, WV and WY.
- Recall number
- Z-0452-2016
- Recall status
- Terminated
- CDRH record ID
- 141982
- FDA classified
- 2015-12-21
- Recall event number
- 72688